ASTM D6978 is the FDA-referenced standard for testing chemotherapy drug permeation through medical gloves. Learn the 9 test drugs, methodology, pass criteria, and which materials qualify.
ASTM D6978 is the standard test method for assessing the resistance of medical gloves to permeation by chemotherapy drugs. It is the primary standard referenced by the FDA for evaluating whether a glove can be labeled as "chemotherapy tested" in the United States. For procurement teams supplying oncology, hematology, and pharmacy compounding departments, understanding this standard is essential for patient and worker safety.
ASTM D6978 (Standard Practice for Assessment of Resistance of Medical Gloves to Permeation by Chemotherapy Drugs) specifies a standardized method for measuring how long a medical glove can resist the permeation of specific oncology drugs. The standard was developed because general chemical permeation standards such as EN 374 do not adequately represent the unique molecular properties and concentrations of chemotherapy agents used in clinical practice.
In the US, the FDA references ASTM D6978 in its 510(k) review process for gloves that carry chemotherapy-related claims. Gloves that pass the standard may be labeled as tested for chemotherapy drug permeation, but the specific drugs tested and breakthrough times must be documented in the Technical Data Sheet (TDS).
ASTM D6978 requires testing against 9 chemotherapy drugs that represent a range of molecular sizes, solubilities, and clinical usage patterns. These drugs were selected because they are among the most commonly handled hazardous drugs in oncology practice.
| # | Drug Name | Class | Clinical Use |
|---|---|---|---|
| 1 | Carmustine (BCNU) | Alkylating agent | Brain tumors, lymphoma |
| 2 | Cisplatin | Platinum compound | Solid tumors, testicular, ovarian |
| 3 | Cyclophosphamide | Alkylating agent | Leukemia, lymphoma, breast |
| 4 | Dacarbazine | Alkylating agent | Melanoma, Hodgkin lymphoma |
| 5 | Doxorubicin HCl | Anthracycline | Breast, leukemia, sarcoma |
| 6 | Etoposide | Topoisomerase inhibitor | Lung, testicular, lymphoma |
| 7 | 5-Fluorouracil (5-FU) | Antimetabolite | Colorectal, breast, gastric |
| 8 | Ifosfamide | Alkylating agent | Sarcoma, testicular, lymphoma |
| 9 | Paclitaxel | Taxane | Breast, ovarian, lung |

The test uses a two-chamber permeation cell. The chemotherapy drug solution is placed in one chamber, and the glove material is clamped between the two chambers as a barrier. The collecting medium in the second chamber is then analyzed at timed intervals to detect whether the drug has permeated through the glove material.
The key performance metric is the permeation rate, measured in micrograms per square centimeter per minute (ug/cm2/min). A glove passes the standard if the permeation rate does not exceed 0.01 ug/cm2/min for each of the 9 drugs over the test period, which typically runs for 240 minutes (4 hours).
| Factor | ASTM D6978 | EN 374 |
|---|---|---|
| Scope | Chemotherapy drug permeation | Industrial chemical permeation |
| Number of test chemicals | 9 oncology drugs | 18 industrial chemicals |
| Test duration | 240 minutes | Typically 480 minutes |
| Pass criterion | Permeation rate <= 0.01 ug/cm2/min | Breakthrough time classification (Type A/B/C) |
| Primary market | United States (FDA) | European Union (MDR) |
| Typical use | Oncology/chemotherapy gloves | Chemical/industrial PPE gloves |
Material performance varies significantly under ASTM D6978 testing:
Q: What is the difference between ASTM D6978 and EN 374?
A: ASTM D6978 specifically tests chemotherapy drug permeation through medical gloves using 9 oncology drugs. EN 374 tests chemical permeation using a broader list of 18 industrial chemicals. They serve different purposes and a glove may meet both standards.
Q: Which chemotherapy drugs are tested under ASTM D6978?
A: The standard requires testing with 9 drugs: carmustine (BCNU), cisplatin, cyclophosphamide, dacarbazine, doxorubicin hydrochloride, etoposide, 5-fluorouracil (5-FU), ifosfamide, and paclitaxel.
Q: Can PVC (vinyl) gloves be used for chemotherapy handling?
A: PVC gloves generally do not pass ASTM D6978 testing for all 9 chemotherapy drugs. Nitrile rubber and natural rubber latex gloves typically perform better. PVC gloves are not recommended for chemotherapy drug handling.
Q: How long must a glove resist permeation under ASTM D6978?
A: The test typically runs for 240 minutes. A glove must demonstrate a permeation rate that does not exceed 0.01 micrograms per square centimeter per minute for each drug tested.
Q: How can I verify that a glove is chemotherapy-tested?
A: Check the Technical Data Sheet (TDS) and Certificate of Conformity (CoC) for explicit reference to ASTM D6978. Be cautious of vague claims like 'chemo-safe' without a specific standard reference.