UDI-DI and EUDAMED for Medical Examination Gloves: EU MDR Labelling, Importer Verification and the 28 May 2027 Certificate Deadline
Jixiang (Xiamen) Medical Technology Co., Ltd.

Since 28 May 2026, registration of economic operators, UDI and device data, notified body certificates and market surveillance in EUDAMED has been mandatory under the EU MDR. For medical examination gloves, that means a valid CE certificate is no longer the whole market-access story: the manufacturer needs an SRN, the device needs a Basic UDI-DI and a UDI/device record, and the label needs the correct UDI carrier. Devices already on the EU market before 28 May 2026 that continue to be supplied face a legacy-registration deadline of 28 November 2026, with notified body certificate back-loading running to 28 May 2027.
The identifiers, in the order they are created
Most documentation errors come from mixing up four different items:
- SRN (Single Registration Number): issued to an economic operator — manufacturer, EU authorised representative or importer — once registered in EUDAMED. A non-EU manufacturer must obtain its own SRN and link its authorised representative; the representative cannot register on the manufacturer's behalf.
- Basic UDI-DI: the device-family identifier that appears in the EU declaration of conformity, technical documentation and certificates. It is not printed on the retail label as a barcode.
- UDI-DI: identifies a specific device and packaging configuration. It is the number submitted to the UDI/device registration module.
- UDI carrier: the machine-readable and human-readable form on the label and higher levels of packaging, produced under a Commission-designated issuing entity such as GS1, HIBCC, ICCBBA or IFA.
Under MDR (EU) 2017/745, UDI carrier requirements were phased by risk class, with Class I devices the last group to come into scope. Non-sterile examination gloves are generally treated as Class I under MDR classification rules, while sterile gloves fall into a higher class requiring notified body involvement — classification must always be confirmed by the manufacturer against Annex VIII rather than assumed from the product name.
The date chain every importer should keep on file
The phased EUDAMED rollout was enabled by Regulation (EU) 2024/1860 and triggered by Commission Decision (EU) 2025/2371, published in the Official Journal on 27 November 2025.
- 28 May 2026: the first four EUDAMED modules became mandatory. New devices must be registered before being placed on the market, and economic operators need an SRN.
- 28 November 2026: legacy and MDR devices first placed on the market before 28 May 2026, and still being supplied afterwards, must be registered. Some authorities describe the cut-off as 27 November 2026 because the legal wording counts twelve months from publication; treating 27 November as the internal target avoids the ambiguity.
- 28 May 2027: notified bodies complete uploading certificate information for devices already registered in EUDAMED.
- Vigilance and post-market surveillance, and clinical investigation modules: rolled out separately once declared functional; confirm the current status with your notified body rather than assuming all six modules are live.
Incomplete or inconsistent registration is not a paperwork inconvenience. It can block customs clearance, trigger marketplace delisting, and leave a CE certificate unable to support lawful placement on the EU market.
An importer verification checklist for glove shipments
Before accepting a 2027 shipment of medical examination gloves, verify the following against EUDAMED and the physical label:
- The manufacturer's SRN and the EU authorised representative's SRN, and that the two are linked to the same legal manufacturer named on the packaging.
- The Basic UDI-DI on the declaration of conformity matches the one on the CE certificate and in the technical documentation.
- A UDI/device record exists for each size and packaging configuration actually shipped, with the expected UDI-DI values.
- The UDI carrier on the dispenser box and shipping carton scans correctly and its human-readable digits match the registered data.
- For products first placed on the market before 28 May 2026, written evidence that legacy registration was completed before the November 2026 deadline.
- Device name, intended purpose, risk class and sterile status agree across the label, the declaration of conformity and the EUDAMED record.
- Your own importer SRN is active and the internal procedure links incoming lots to the EUDAMED record version.
Ask for screenshots or an export of the EUDAMED record with the shipment documents. Reconstructing this evidence during a customs hold is slow and expensive.
FAQ
Q: Is a CE certificate alone enough to import medical gloves into the EU in 2027?
A: No. The manufacturer also needs an SRN, a Basic UDI-DI, a UDI/device registration in EUDAMED and a compliant UDI carrier on the label. A CE certificate without the corresponding registration can still block a shipment.
Q: What is the difference between Basic UDI-DI and UDI-DI?
A: Basic UDI-DI identifies the device family and appears in the declaration of conformity and certificates. UDI-DI identifies a specific device and packaging configuration and is the identifier submitted to EUDAMED and printed on the label.
Q: Can an EU authorised representative register a non-EU manufacturer's SRN?
A: No. A non-EU manufacturer must apply for its own SRN in its own name and link the representative's SRN. Check who holds each number before accepting a compliance pack.
Source note: EU MDR (EU) 2017/745, Articles 27 and 29 and Annex VI Part C; Regulation (EU) 2024/1860 on the phased rollout of EUDAMED; Commission Decision (EU) 2025/2371 (OJ, 27 November 2025); European Commission EUDAMED Information Centre and MDR/IVDR UDI guidance. Deadline wording differs by one day between some authority publications; verify the consolidated timeline with your notified body.