ISO 11193 is the international standard specification for single-use medical examination gloves. It has two parts: ISO 11193-1 covers gloves made from rubber latex or rubber solution (natural rubber latex, nitrile rubber, polychloroprene, styrene-butadiene rubber, and thermoplastic elastomers), and ISO 11193-2 covers gloves made from poly(vinyl chloride), PVC. The current edition of Part 1 is ISO 11193-1:2020. When a glove carries ISO 11193 compliance, it means the lot has been tested for freedom from holes, dimensions, physical properties (force at break and elongation), and biocompatibility.
ISO 11193-1:2020 (third edition, published August 2020) specifies requirements for packaged sterile or bulk non-sterile rubber gloves used in medical examinations and diagnostic or therapeutic procedures, and for handling contaminated medical materials. Gloves are classified by type and finish:
| Part | Material | Typical glove | Current edition in wide use |
|---|---|---|---|
| ISO 11193-1 | Rubber latex / rubber solution (natural rubber, nitrile, polychloroprene, SBR, TPE) | Medical examination gloves | ISO 11193-1:2020 |
| ISO 11193-2 | Poly(vinyl chloride), PVC | Vinyl examination gloves | ISO 11193-2 |
ISO 11193-1 sets requirements across five areas, with test methods drawn from other standards:
| Attribute | Requirement (typical) | Test method |
|---|---|---|
| Length | Min 220-230 mm depending on size | ISO 11193-1 dimensional requirements |
| Thickness | Min 0.05-0.08 mm at palm/finger (varies by finish) | ISO 4648 |
| Freedom from holes | AQL 1.5 (water leak test, ISO 2859-1 sampling) | Water leak test |
| Force at break | >= 7 N before aging; >= 6 N after aging (median) | ISO 37 |
| Elongation at break | >= 500% before aging; >= 400% after aging | ISO 37 |
| Aging | 7 days at 70 degrees C | ISO 188 |
| Powder residue (powder-free) | Maximum powder per glove, if labeled powder-free | Defined in standard |
| Biocompatibility | Cytotoxicity, sensitization, irritation | ISO 10993 series |
Note: values above reflect typical published conformance data from major manufacturers; always verify the exact figures in the supplier batch test report. AQL stands for Acceptable Quality Limit - for medical examination gloves the hole tolerance is AQL 1.5 (maximum 1.5% of a sampled lot may leak).
Buyers often see ISO 11193, EN 455, and ASTM D6319 on the same datasheet. They test similar properties but express them differently:
| Property | ISO 11193-1 | EN 455-2 | ASTM D6319 |
|---|---|---|---|
| Market focus | International / GCC (GSO ISO 11193-1) | EU | US (FDA 510(k)) |
| Freedom from holes | AQL 1.5 | AQL 1.5 (EN 455-1) | AQL 1.5 |
| Force at break (exam gloves) | >= 7 N (before aging), >= 6 N (after) | >= 6.0 N | Not used (uses MPa) |
| Tensile strength | Not specified as MPa | Not specified as MPa | >= 14 MPa (before and after aging) |
| Elongation | >= 500% / >= 400% | Not specified | >= 500% / >= 400% |
| Aging protocol | 7 days at 70 degrees C | 7 days at 70 degrees C | 7 days at 70 degrees C (per ASTM D7160) |
| Biological evaluation | ISO 10993 series | EN 455-3 | ISO 10993 series |
The key insight: EN 455-2 and ISO 11193 express strength as force at break in Newtons, while ASTM D6319 expresses it as tensile strength in MPa. Both are valid; they measure different quantities (absolute force vs force per unit cross-section).

Data sources: ISO 11193-1:2020 (ISO.org); ISO 11193-2; GSO ISO 11193-1:2020; EN 455-1/2 series; ASTM D6319-19; ISO 2859-1; ISO 37; ISO 188; ISO 10993 series. Values cross-checked against published manufacturer conformance data. Information current as of August 2026.
Q: Is ISO 11193 the same as EN 455 or ASTM D6319?
A: No, but they cover similar ground. ISO 11193 is the international standard for single-use examination gloves; EN 455 is the EU harmonized standard (Part 1 holes, Part 2 physical properties, Part 3 biological evaluation, Part 4 shelf life); ASTM D6319 is the US standard for nitrile examination gloves linked to FDA 510(k) clearance. Many export gloves are certified to two or three of these. Key differences: EN 455-2 and ISO 11193-1 report strength as force at break in Newtons, while ASTM D6319 reports tensile strength in MPa.
Q: Does ISO 11193 apply to PVC (vinyl) gloves?
A: Yes. ISO 11193-2 is the part that specifies requirements for single-use examination gloves made from poly(vinyl chloride). Part 1 covers rubber materials (natural rubber latex, nitrile rubber, polychloroprene, SBR, TPE). When a datasheet simply says ISO 11193, check which part is referenced - Part 1 for rubber, Part 2 for PVC.
Q: What AQL does ISO 11193 require for pinholes?
A: ISO 11193-1 requires an Acceptable Quality Limit of 1.5 for freedom from holes when tested with the water leak test and sampled per ISO 2859-1. This means no more than 1.5% of the sampled lot may leak. AQL 1.5 is the same threshold used by EN 455-1 and ASTM D6319 for medical examination gloves.
Q: How should a buyer verify ISO 11193 compliance?
A: Ask the supplier for the certificate and, more importantly, the batch test report showing the standard and edition, the batch code, the water-leak AQL result, force at break and elongation before and after aging, powder residue (if powder-free), and ISO 10993 biocompatibility data. The report must cover the specific batch being shipped, not a historical sample. You can also verify that the test laboratory is accredited (e.g., ISO/IEC 17025).
Q: Can a glove be certified to ISO 11193 and EN 455 at the same time?
A: Yes, this is common practice. Export factories typically certify a single product to ISO 11193-1 (international/GCC), EN 455 (EU), and ASTM D6319 (US) so the same glove can be sold across markets. Because the test methods for holes and aging are harmonized, dual certification usually requires little extra testing beyond the physical property differences (force at break vs tensile strength).